Additional information added to h3, h6 and h10.H10: the device was evaluated on site by the qualified technician.It was confirmed the excess ultrafiltration of 500 ml during treatment.To resolve the issues, the qualified technician replaced the ultrafiltration unit, bypass valve (byva), taration valve (tava), direct valve (diva), rinse valve (riva), zero (uf) valve (zeva) and degassing level sensor.A service history review this machine involved in this complaint determined that no similar events occurred on this machine.At the moment no specific actions will be initiated; a data monitoring system is maintained for the events reported from the field as well as for quality data from internal processes.Should additional relevant information become available, a supplemental report will be submitted.
|