Model Number DBP-200SOLID145
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Return of the device for analysis is anticipated.If the device is received for analysis, a supplemental report will be submitted when the device analysis is completed.Csi id: (b)(4).
 
Event or Problem Description
Following treatment in the superficial femoral artery, the diamondback 360 peripheral orbital atherectomy device (oad) became stuck in the vessel upon removal.The oad was turned on and pulled retrograde.The patient was stable with no consequences.
 
Additional Manufacturer Narrative
Results code: 4247 - suggested code is biological material present on device.A section of driveshaft 16.1cm long was returned to csi without the oad.Scanning electron microscopic testing was performed and confirmed the driveshaft was cut.Biological material accumulation was observed on the distal edge of the crown which was not considered to be a contributing factor in the reported complaint.No damage or abnormalities were observed on the driveshaft or crown that would have contributed to the reported event.As the non-csi sheath was not returned, it could not be determined if it was damaged or contributed to the reported complaint.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15463313
Report Number3004742232-2022-00225
Device Sequence Number20544582
Product Code MCW
UDI-Device Identifier10850000491202
UDI-Public(01)10850000491202(17)221031(10)349975-1
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberDBP-200SOLID145
Device Catalogue Number7-10057-04
Device Lot Number349975-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-