| Model Number |
03.614.017 |
| Medical Device Problem Code |
Compatibility Problem (2960)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, during o-t fusion (cervical spine) for cervical myelopathy, the screw in question could not be attached to the two screwdrivers in question.The operation was completed successfully using another synapse screwdriver.There was a delay of approximately 30 minutes.No further information is available.This report is for a threaded holding sleeve for polyaxial screws.This is report 2 of 7 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: a review of the device history records was performed for the finished device lot number and no nc were found that impacted the complaint condition.All necessary actions to ensure the final product quality have been taken and documented in the appropriate quality system.The final quality release criteria were met before this batch was released for distribution.Part number: 03.614.017; lot number: t150587; manufacturing site: tuttlingen; release to warehouse date: 04-apr-2017.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects on the surface of hold-sl, p/n: 03.614.017.A dimensional inspection was unable to be performed due to not being relevant to the complaint condition.A functional test was performed.The outer sleeve was assembled to the returned hexagonal screwdriver shaft cross pinned, p/n: 03.614.039, and the returned hold-sl, p/n: 03.614.017.After that, the returned cancellousscr synapse ø4 l26 f/r ø4 tan, p/n: 04.615.126s, was threaded to the assembly.The screw was successfully assembled.The complaint condition was not replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the observed condition of the hold-sl, p/n: 03.614.017 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|