• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH SUCTION YANKAUER CLEAR BULB TIP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH SUCTION YANKAUER CLEAR BULB TIP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number K86
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
Customer reports: during a total hip replacement the bulbous tip of the yankauer suction handle broke off in the femoral canal and was unable to be removed.
 
Additional Manufacturer Narrative
A review of the device history records (dhr) yielded no discrepancies within the manufacture of the affected lot number reported.All quality assurance testing performed during manufacturing was acceptable.A sample analysis could not be performed since no photo samples nor actual sample device was received for evaluation.The root cause could not be determined without a sample to evaluate.Current controls include a 100% visual inspection by production personnel according to the standard work instruction for yankauer bulk products to inspect for damaged and broken components.In addition, a quality control sampling inspection is performed to ensure that product complies with the visual inspection of damage/broken components.In-process controls such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant are in place to prevent nonconforming product from leaving the manufacturing operations.Based on the assessment, no further actions are being pursued as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUCTION YANKAUER CLEAR BULB TIP
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15464181
Report Number9612030-2022-03396
Device Sequence Number6902732
Product Code GCX
UDI-Device Identifier00885380030157
UDI-Public00885380030157
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberK86
Device Catalogue NumberK86
Device Lot Number2208302664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
MW0300060000-2022-8006
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
Patient SexMale
-
-