Model Number AMSCO 400 STEAM STERILIZER
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician inspected the sterilizer and found that the pressure gauge to their amsco 400 sterilizer required replacement.The pressure gauge was not installed by steris and had been installed by a third-party service provider.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that water was leaking from their sterilizer.No report of injury.
 
Additional Manufacturer Narrative
The steris service technician ran all test cycles for the sterilizer and confirmed the unit to be operating properly; no issues were noted.The reported event may be attributed to the third-party installation of the pressure gauge.The user facility is currently closed and stated they will contact steris for re-install and start-up of the amsco 400 sterilizer when they are ready to place the unit back in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO 400 STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15464189
Report Number3005899764-2022-00046
Device Sequence Number6902735
Product Code FLE
UDI-Device Identifier00724995209254
UDI-Public00724995209254
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAMSCO 400 STEAM STERILIZER
Device Catalogue NumberSR0101110311
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer08/22/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-