MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS M.SA20 U.CONTROL RESERV.; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS M.SA20 U.CONTROL RESERV.; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX434T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 08/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fx434t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the shunt system was believed to be operated in underdrainage and was difficult to adjust.Reportedly, following the primary surgery, the patient showed no improvement in hydrocephalus, so pressure changes were performed several times.After that, when the doctor tried to change the pressure, the pressure could not be changed.The doctor could push the reservoir, but felt resistance.The doctor believes that the protein component may have clogged the valve.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 4 years, 3 months; gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: cut in the kink protection between the valves.Permeability test: the test showed that the progav 2.0 shunt system is permeable.Computer control test: the computer controlled test showed the opening pressure of the progav2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 0.44 cmh2o.This is not within the specified tolerance of 8 cmh2o ± 3 cmh2o.An applied pressure of 8 cmh2o, with the device in the horizontal position is expected to have a resultant pressure of 8 cmh2o ± 3 cmh2o.According to our result, we can determine a presence of an accelerated outflow.Additionally, the shunt assistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 20 cmh2o in the vertical position, a pressure of 20 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shunt assistant had a pressure of 19.23 cmh2o.This is within the specified tolerance of 20 cmh2o +4/-2 cmh2o.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.The shunt assistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated shunt system was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valves were finally opened with a special tool.After dismantling of the valves, deposits were found in progav 2.0.In the shunt assistant were no visible deposits.Results: based on our investigation results, we have determined that there was an accelerated outflow in the progav 2.0 valve at the time of our investigation.The detected deposits could be the cause of the accelerated outflow.Existing deposits in the csf can temporarily impair the function of the valve and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS M.SA20 U.CONTROL RESERV.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15464898
Report Number3004721439-2022-00286
Device Sequence Number6897027
Product Code JXG
UDI-Device Identifier04041906137328
UDI-Public4041906137328
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX434T
Device Catalogue NumberFX434T
Device Lot Number20056727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-