• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYSTEM WITH SPRUNG RESERVOIR; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYSTEM WITH SPRUNG RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX424T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: no abnormalies detected.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav 2.0 was tested and is not adjustable throughout the normal range.Klick force and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.(see section adjustment test).Computer controlled test: the results show that the valve is not operating within the acceptable tolerance in the hoirzontal position.Internal inspection: after dismantling of the valve, deposits were found in progav 2.0.Result: based on our investigation, we confirm that the valve was not adjustable and showed an accelerated outflow at the time of our investigation.This is likely due to the deposits observed inside the valves.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 (part # fx424t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be operated in overdrainage and had adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 2 years, 7 months.Height: 94 centimeters (cm).Weight: 11 kilograms (kg).Gender: male.Same event as medwatch# 3004721439-2022-00281.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYSTEM WITH SPRUNG RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15465097
Report Number3004721439-2022-00280
Device Sequence Number6897050
Product Code JXG
UDI-Device Identifier04041906136949
UDI-Public4041906136949
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2026
Device Model NumberFX424T
Device Catalogue NumberFX424T
Device Lot Number20053923
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2022
Initial Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 YR
Patient SexMale
Patient Weight11 KG
-
-