| Model Number |
AMSCO 600 STEAM STERILIZER |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician inspected the amsco 600 sterilizer and found the water hose to their unit to be damaged.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the water hose to their amsco 600 sterilizer disconnected subsequently causing water to leak out onto the floor.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The technician replaced the hose, tested the unit and confirmed it to be operational.The sterilizer was returned to service and no additional issues have been reported.
|
| |
|