Model Number VTICM5_13.2
Medical Device Problem Codes Off-Label Use (1494); Adverse Event Without Identified Device or Use Problem (2993); Unintended Movement (3026)
Health Effect - Clinical Codes Cataract (1766); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon inserted and removed a 13.2mm vticm5_13.2 implantable collamer lens, -09.50/+3.0/093 (sphere/cylinder/axis), within the same surgery due to the lens flipped in the eye.The backup lens was implanted.Additional information has been requested but none has been forthcoming.
 
Additional Manufacturer Narrative
Additional data: h6: health impact: additional surgery: 4625: cataract surgery, iol implanted.H6: device code: 1494: this lens model is contraindicated for patients with age more than that of 45 years.Corrected data: b5: the reporter indicated the surgeon inserted and removed a 13.2mm vticm5_13.2 implantable collamer lens, -09.50/+3.0/093 (sphere/cylinder/axis), within the same surgery due to the development of an anterior subcapsular/cortical cataract.Cataract surgery (phacoemulsification) was performed the next day and toric iol was implanted.The patient is doing well.The cause of the event was due to user error (the eye was too soft and patient squeezed when the incision for icl was made, cutting the anterior capsule, which induced the cataract).Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key15465183
Report Number2023826-2022-03166
Device Sequence Number6897060
Product Code QCB
UDI-Device Identifier00841542116541
UDI-Public00841542116541
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberVTICM5_13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL AND LOT # UNK; FOAM TIP PLUNGER MODEL FTP - LOT # 1581842; INJECTOR MODEL AND LOT # UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexMale
-
-