GYRUS ACMI, INC. HALO PKS CUTTING FORCEPS,5MM/33CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number HACF0533
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
While onsite at the facility, the olympus account manager observed the hacf0533 halo cutting forceps did not activate during a procedure.The procedure was completed with an acceptable outcome.The procedure and type were unknown.There was no report of patient harm as a result of the event.This was 1 of 3 cutting forceps that failed during the procedure.Related reports include the following patient identifiers:(b)(6).
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Please see updates to d9, h3, h4, h6 and h10.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it¿s likely the reported event (halo cutting forceps did not activate) was due to insufficient generator output time, and/or having the generator not connected to the device settings.During evaluation, the device was tested in both ports (a and b) and was able to operate and coagulate as intended.The root cause of this event was unable to be identified.The following is included in the instructions for use: ¿warning: insufficient generator output time or generator output level may not provide the operator with the desired level of coagulation." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALO PKS CUTTING FORCEPS,5MM/33CM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15465312
Report Number3011050570-2022-00115
Device Sequence Number12473291
Product Code GEI
UDI-Device Identifier00821925036390
UDI-Public00821925036390
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K100896
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHACF0533
Device Lot NumberFR101646
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-