Catalog Number 114364 |
Device Problem
Filtration Problem (2941)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/30/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that an excess fluid loss (amount not specified) occurred during treatment with an ak 96 machine.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Initial reporter first name: initial reporter last name: initial reporter facility name: initial reporter address: should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information added to h6 and h10.H10: the device was evaluated on site by a qualified technician.To resolve the issue, the qualified technician calibrated the uf cell and the machine was running normally.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|