MAUDE Adverse Event Report: BERLIN HEART GMBH BERLIN HEART IKUS STATIONARY DRIVING UNIT; VENTRICULAR ASSIST DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BERLIN HEART GMBH BERLIN HEART IKUS STATIONARY DRIVING UNIT; VENTRICULAR ASSIST DEVICE Back to Search Results
Model Number D03I-111
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Ischemia (1942)
Date of Event 08/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The site initially contacted clinical affairs (ca) to report that the" pump stopped working", and the patient was being supported using the hand pump.Ca heard that the ikus was audibly alarming in the background.The site could not specify if the blood pump or the ikus unit had stopped working.Per the site, the patient was receiving cpr while being hand pumped; however, the site caller was unable to verify if the blood pump was achieving complete fill and ejection during hand pumping.Prior to initiation of hand pumping and cpr, the nurse entered the patient's room and noted the patient was "posturing, seizing and rigid," and thus, a code was called.The patient was supported via hand pump for approximately 10-15 minutes, while the backup ikus was prepared.The backup ikus was programmed to the patients' current vad settings and connected to the patient.The site reported that, once the backup ikus was connected to the patient, the blood pump was functioning as expected with full fill and full ejection.The ikus unit in question was restarted to collect the log files which were downloaded and sent to bhi for review by the service engineers.Of note, prior to downloading the log files, the ikus was turned off to stop it from alarming.When the unit was restarted to collect the log files, the machine passed the self test and started without incident.The ikus unit was collected on complaint.
 
Additional Manufacturer Narrative
When the ikus was received from the site, it was started, passed the self test and functioned as expected without any incident.The product at the manufactures request, was sent to the manufacturer, berlin heart gmbh for evaluation.Currently, evaluation of the ikus is being planned, however the device is being held in quarantine as berlin heart has received legal notice to preserve the device in the condition that it was received.The manufacturer will proceed with the evaluation once we have permission to do so.According to the clinic patient was withdrawn from support on (b)(6) 2022 due to an anoxic brain injury.A detailed evaluation report will be provided as soon as it is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BERLIN HEART IKUS STATIONARY DRIVING UNIT
Common Device Name
VENTRICULAR ASSIST DEVICE
Manufacturer (Section D)
BERLIN HEART GMBH
wiesenweg 10
berlin, deu 12247
GM  12247
MDR Report Key15465402
Report Number3008454189-2022-00012
Device Sequence Number17853185
Product Code DSQ
UDI-Device Identifier04260090040225
UDI-Public04260090040225
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 09/21/2022
Report Date (Section F) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberD03I-111
Device Catalogue NumberD03I-111
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2022
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/20/2022
Device Age14 YR
Event Location Hospital
Date Report to Manufacturer08/20/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/22/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age4 YR
Patient SexMale
Patient Weight18 KG
Date Report Sent to FDA09/21/2022
-
-