| Model Number |
VICM5_13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923); Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Type of investigation: 4110 - lens work order search: one similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vicm5_13.2; -07.50 diopter implantable collamer lens into the patient's left eye (os).The patient experienced low vaulting.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5 - the reporter indicated that the surgeon implanted a 13.2mm vicm5_13.2; -07.50 diopter implantable collamer lens into the patient's right eye (od) on (b)(6) 2022.The patient experienced low vaulting; glare; lens dislocation/subluxation.On (b)(6) 2022 the lens was explanted and on this same day a replacement lens of longer length was implanted and this resolved the problem.The cause of the event was reported as unknown and noted "low vault - could have been mis-measured or abnormal ratio of diameter to white to white.Claim#: (b)(4).
|
| |
|