• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO REFLECT METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO REFLECT METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4859765
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Nausea (1970); Paresis (1998); Diaphoresis (2452); Confusion/ Disorientation (2553)
Event Date 08/22/2022
Event Type  Injury  
Event Description
On (b)(6) 2022, the reporter of the lay user/patient contacted lifescan (lfs) usa, alleging that the patient¿s onetouch verio flex meter was reading inaccurately high compared to his feelings and/or normal readings.This complaint was classified based on information obtained from the customer care agent (cca) during the initial call.The reporter alleged that the product issue began on (b)(6) 2022 at 5 pm.The reporter claimed that the patient obtained blood glucose readings of ¿587 and 487 mg/dl¿ on the subject meter.The patient manages his diabetes with an unspecified insulin (self-adjusting dose, twice a day) and the reporter stated that the patient increased his insulin intake to 10 units at 5:10 pm in response to the high readings.The reporter informed the cca that at 7 pm that same day, the patient started to develop symptoms of ¿confusion, nausea, can¿t get up and sweating¿.The reporter claimed that the patient was treated with food and/or drink, provided by another person.At 7:30 pm, after treatment, a blood glucose reading of ¿120 mg/dl¿ was obtained on another unspecified meter.The reporter indicated that the patient felt better a few minutes after treatment.During troubleshooting, the cca noted that the subject meter was set to the correct unit of measure.The cca confirmed that the patient¿s test strips had been stored correctly, were within expiry date and the test strip vial was not cracked or broken.The patient did not have control solution available to test the subject meter.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed symptoms suggestive of a serious injury adverse event after taking an increased dose of insulin based on alleged inaccurate high results obtained with the subject meter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO REFLECT METER
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key15466378
MDR Text Key300407042
Report Number2939301-2022-03103
Device Sequence Number1
Product Code NBW
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number4859765
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/26/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/22/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
-
-