Model Number 21-2770-24
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The water cannot flow through the gripper product, the fluid stops before pass the base.The water could not flow through the gripper product, the fluid stopped before pass the base.Manufacturing controls are in place to help prevent occurrence of the confirmed issue; however, a correction action preventive action (capa) was issued, in order to determine the root cause and implement proper corrective actions if applicable upon completion of the investigation.Operator of device is unknown.No information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that when performing priming at a pre-use check, the customer noticed medical fluid did not pass through the product due an occlusion in it.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER PLUS NEEDLES
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
sapporo-city, hokkaido
minneapolis, MN 55442
MDR Report Key15466980
Report Number3012307300-2022-20144
Device Sequence Number20585892
Product Code FPA
UDI-Device Identifier10610586025631
UDI-Public10610586025631
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2024
Device Model Number21-2770-24
Device Catalogue Number21-2770-24
Device Lot Number3841564
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Initial Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-