• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ENDOBRONCHIAL TUBES; TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ENDOBRONCHIAL TUBES; TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION Back to Search Results
Model Number 198-32L
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Five photos and one sample was returned for evaluation.The sample was received in used condition within its original package.Visual and functional testing were performed.Visual inspection determined that no defects were observed.Functional testing found that there was resistance inflating and deflating the cuff; occlusion failure mode was confirmed.The occurrence of this failure condition could be caused by method not followed (excess solvent thf at joint) at manufacture.Operator of device was unknown.No information has been received to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that when inflating and deflating the tracheal cuff at a pre-use check, the customer felt extreme resistance through it.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ENDOBRONCHIAL TUBES
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15467418
Report Number3012307300-2022-20149
Device Sequence Number20553352
Product Code CBI
UDI-Device Identifier15019315058608
UDI-Public15019315058608
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K953483
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2023
Device Model Number198-32L
Device Catalogue Number198-32L
Device Lot Number3955980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Initial Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-