Model Number INC-11597-132
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacture.
 
Event or Problem Description
It was reported during aspiration with the pump when the subject catheter was retracted, there was severe resistance.After removing the catheter from the patient, the tip of the device was found fractured.The device was replaced and the procedure was completed successfully with no clinical consequences to the patient.
 
Event or Problem Description
It was reported during aspiration with the pump when the subject catheter was retracted, there was severe resistance.After removing the catheter from the patient, the tip of the device was found fractured.The device was replaced and the procedure was completed successfully with no clinical consequences to the patient.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection the catheter shaft was seen to be stretched from 6cm to 10cm from the distal tip, was seen to be flat in several areas, and the catheter tip was seen to be flat.Functional inspection was not carried out due to the condition of the returned device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event was not confirmed during analysis as no breaks were present anywhere on the catheter and functional inspection was not carried out due to condition of device.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information provided by the customer indicated that there was no damage noted to the packaging prior to preparation of the device, there were no anomalies noted to the device after removal from the packaging or prior to preparation and the device was prepared as per the dfu.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefore, an assignable cause of procedural factors will assigned to the analyzed defect of the catheter shaft stretched, catheter shaft flat/crushed and catheter tip flat/crushed.An assignable cause of 'not confirmed' was assigned to the reported defect of the catheter tip broken/fractured during use and the catheter shaft difficulty removing/withdrawing as the issue included a returned product review (visual, physical, and/or performance testing) which showed no evidence of either the alleged issue(s) or any defect which could have contributed to the event.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS VECTA 074 CATH 132CM - US
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork CA NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork CA NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15468664
Report Number3008853977-2022-00046
Device Sequence Number19914282
Product Code NRY
UDI-Device Identifier07613327459265
UDI-Public07613327459265
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2024
Device Model NumberINC-11597-132
Device Catalogue NumberINC-11597-132
Device Lot Number17188-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/02/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date09/22/2022
Supplement Report FDA Received Date11/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PUMP (UNKNOWN)
Patient SexUnknown
-
-