COVIDIEN PROTEGE RX CAROTID STENT SYSTEM; STENT, CAROTID Back to Search Results
Model Number SECX-8-6-40-135
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
Physician was attempting to use a protégé rx carotid stent during procedure.No damage noted to packaging.No issues noted when removing the device from the hoop/tray.Device prepped per the ifu, with no issues identified.Lesion was not pre-dilated.Device did pass through a previously-deployed stent.No resistance encountered when advancing the device.No excessive force used.It is reported that the device could not be deployed at the target lesion during the procedure.After that, the procedure was completed without any problems using the same products.No intervention required to remove device.Procedure completed with different product of the same make.No patient injury reported.When the stent was returned it was noted that the stent was partially exposed.
 
Additional Manufacturer Narrative
Product analysis the device was returned with the touhy-borst loose.The device was returned with approx.3mm of the stent exposed from the outside tip of the outer assembly the stent confirmed as 60mm a 20cc water filled syringe was used to flush the device without issue, on both annual spaces and inner lumen a 0.014¿ guidewire was loaded through the guidewire lumen without issue, and the device was loaded into a deployment fixture the stent did not deploy below the 3lb pressure limit the stent could not be deployed with a maximum peak force of 3.50lbs medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE RX CAROTID STENT SYSTEM
Common Device Name
STENT, CAROTID
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15468930
Report Number2183870-2022-00312
Device Sequence Number11471134
Product Code NIM
UDI-Device Identifier00821684038703
UDI-Public00821684038703
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P060001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2023
Device Model NumberSECX-8-6-40-135
Device Catalogue NumberSECX-8-6-40-135
Device Lot NumberB256780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-