| Model Number |
67SP040 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Operator of device is patient/consumer.Initial reporter also sent report to fda is unknown.No information has been provided to date.No dhr review was performed.A product sample was received for evaluation.Visual and functional testing were performed.Sample 1, the cuff wasn't inflating uniformly.The root cause of the reported issue was found to be cuff symmetry test of representative sample per lot by quality inspectors.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that cuff would not inflate and the water would not go down the cuff.No patient injury was reported.
|
| |
|