SYNTHES PRODUKTIONS GMBH 3.5MM TI ROD 120MM; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
|
Back to Search Results |
|
| Model Number |
498.125 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Device report from brazil reports an event as follows: it was reported that on (b)(6) 2022, imaging exams confirmed a broken rod.Patient originally underwent spinal cord decompression with cervical spine arthrodesis on (b)(6) 2017, for surgical treatment of basilar intussusception; posterior fossa craniectomy + occipital cervical fixation (arnoldchiari syndrome).Postoperatively, the patient complained of pain and lethargy.This report is for an unknown rod.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: patient id also reported as (b)(6).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.This report is for an unknown rod/unknown lot.Part and lot number are unknown; udi number is unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the unk - rods was broken in several parts.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for unk - rods.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2: additional procodes: nkb, mnh, mni, kwq, kwp.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: device history lot part # 498.125, lot # h109765, manufacturing site: werk selzach, release to warehouse date: 30 aug2016, supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 3.5mm ti rod 120mm was broken in two fragments, both were returned for evaluation.The edges of the broken ends do not present any signs of material issue.Since the surgery was performed in 2017 and the complaint condition was identified in 2022, the breakage of the rod is most likely due to fatigue, result of a non-union fracture followed by continuous micromotion of the device until failure.A dimensional inspection for the 3.5mm ti rod 120mm was unable to be performed due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 3.5mm ti rod 120mm would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|