| Catalog Number |
8065752088 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported there was no power to sculpt the nucleus during a cataract surgery (with iol implant).The surgery was completed after the cassette and the handpiece were replaced.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The customer reported the handpiece did not have power.The handpiece is electrically connected to the console when in use, the cassette contains no electrical conduit that would power the handpiece.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Current tracking indicates no adverse trend for this lot for this event.In-process controls are established to ensure each final assembled cassette is verified that all required tests have been performed and all acceptance criteria are met.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and no obvious defects were observed that would have contributed to the reported event.A calibrated console was used to test the sample and the fluidics management system (fms) cassette primed and tuned with the ultrasonic handpiece successfully and could achieve maximum vacuum.No system message was generated, no fluid or air leaks, and no cracks were observed on the connectors that would have contributed to the reported event.The irrigation and aspiration flow rates were measured and found to be within specifications.Fluid flowed from the balance salt solution (bss) bottle to the irrigation and aspiration manifolds and continuously to the cassette housing.No occlusion or obstruction was identified during inspection and functional testing.The root cause of the customer's complaint could not be established as the returned fluidics management system (fms) cassette was evaluated and functioned per specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Based on our current tracking, there are no adverse trends for this reported complaint and lot.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|