| Model Number |
AED PLUS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was verified and attributed to the integrated circuit on the main board.The main board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|