MAUDE Adverse Event Report: TIDI PRODUCTS, LLC SITTER ELITE; FALL PREVENTION ALARM/SENSOR COMBINATION ATTACHED OR UNATTACHED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TIDI PRODUCTS, LLC SITTER ELITE; FALL PREVENTION ALARM/SENSOR COMBINATION ATTACHED OR UNATTACHED Back to Search Results
Model Number 8345
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 bd1 the customer contacted us via phone call.Customer states that the alarm will not trigger nurse call system at the nurse station.Customer said they tired a new unit with where this alarm was and it worked.No apparent damage to the unit on the nc port or battery compartment.Sn (b)(4).
 
Additional Manufacturer Narrative
Visual findings observed the nurse call receptacle pin 3 inside was bent down.The warning label is peeling off from top edges.Site stickers attached at the lower end of the unit.Evaluation found the unit that it did not send a signal to activate the indicator light at the nurse call test fixture due to a bent pin 3 inside the nurse call receptacle.In addition, due to the bent pin 3, the mode button did not go into voice only and mute modes when the nurse call cable was plugged into the nurse call receptacle.The unit passed all other functional testing.If the device should fail to send a signal to the nurse call station, it may not alert the caregiver of a patient exit.Being that the unit is already more than two years old since it was manufactured, it is unlikely that a manufacturing non-conformity contributed to the bent pins inside the nurse call receptacle, and the likely cause of the event is normal wear and tear.At this time, there is no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use (ifu) were reviewed and determined to provide adequate instructions and warning for the safe and effective use of the device.The ifu states, to reduce the risk of serious injury or death, test the alarm and sensor for proper operation prior to putting in service with a patient, and each time before leaving the patient unattended.If the alarm and/or sensor do not function properly, remove the alarm and sensor from service and replace them with a properly functioning alarm and/or sensor.Do not use the alarm, sensor or magnet if it does not activate each time weight is removed from the sensor, the chair belt sensor is unfastened, or magnet is removed from face plate.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITTER ELITE
Common Device Name
FALL PREVENTION ALARM/SENSOR COMBINATION ATTACHED OR UNATTACHED
Manufacturer (Section D)
TIDI PRODUCTS, LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key15470317
Report Number2182318-2022-00112
Device Sequence Number4468634
Product Code PJO
UDI-Device Identifier10190676002082
UDI-Public10190676002082
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8345
Device Catalogue Number8345
Device Lot Number2012KF0718
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Initial Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1