• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.Upon inspection testing of the returned device, the user's request was confirmed.It was discovered, the device failed the water dunk test due to a hole at the bending section which was punctured due to excessive force.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus, the uretero-reno videoscope failed the leak test multiple times during reprocessing.There were no reports of patient harm associated with this event.
 
Event Description
This medical device report was submitted for the damage found at estimation.The a-rubber was punctured by a broken bending section.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, to update d8, and make corrections to b5 and h6 coding.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the user applied excessive stress to the bending section which resulted in the damage to the bending tube.The event can be detected and prevented by following the instructions for use (ifu) which state: "-precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.-inspection of the endoscope : inspect the external surface of the entire insertion section including the bending section and the distal end for any irregularities such as dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, and protruding objects." "-do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.-do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15470495
MDR Text Key306417664
Report Number9610595-2022-02105
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403385
UDI-Public04953170403385
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-