• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL EZ-IO INTRAOSSEOUS INFUSION SYSTEM - DRIVERS; NEEDLE, HYPODERMIC, SINGLE LU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL EZ-IO INTRAOSSEOUS INFUSION SYSTEM - DRIVERS; NEEDLE, HYPODERMIC, SINGLE LU Back to Search Results
Model Number IPN915667
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
Ez-io power driver failed to operate.The light would initially illuminate green for less than a second and then the device would die out.The device no longer has any light turn on and does not work.There was never a red light that illuminated on the device.Another driver was obtained.There was no patient injury or consequence.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.Two ifus were referenced for this investigation.The ez-io driver ifu states, "as with any emergency medical device carrying a backup is strongly advised protocol", "ez-io power driver led will blink red when the trigger is activated and has only 10% of battery life remaining", and "purchase and replace the ez-io power driver when the red led begins blinking"."when storing the vascular access pak (vap), remove the trigger guard to prevent accidental activation of the ez-io power driver".The ez-io needle ifu "squeeze trigger and apply gentle, steady pressure.Important: do not use excessive force.Note: if ez-io power driver stalls and ez-io needle set will not penetrate the bone, operator may be applying too much downward pressure to penetrate bone.".A review of the device history record found that the driver passed all the release criteria.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Manufacturer will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Ez-io power driver failed to operate.The light would initially illuminate green for less than a second and then the device would die out.The device no longer has any light turn on and does not work.There was never a red light that illuminated on the device.Another driver was obtained.There was no patient injury or consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ-IO INTRAOSSEOUS INFUSION SYSTEM - DRIVERS
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LU
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15470725
Report Number3011137372-2022-00186
Device Sequence Number11842961
Product Code FMI
UDI-Device Identifier10816000013813
UDI-Public10816000013813
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K180395
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915667
Device Catalogue Number9058
Device Lot NumberK55075
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/18/2022
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-