AOMORI OLYMPUS CO., LTD. SONOSURG CURVED SCISSORS,HF,PISTOL GRIP,5MM X 34CM; ULTRASONIC SURGICAL DEVICE
|
Back to Search Results |
|
| Model Number |
T3905 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation.Upon inspection testing of the returned device, the user's request was confirmed.It was discovered, the probe was broken at approximately 16 mm from the tip.Upon checking the fracture surface, a black discoloration was found on the fracture surface of the probe and cracks were spreading from the black discoloration.There were no flaws around the probe break point.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
An olympus representative reported to olympus, on behalf of the customer, the probe broke while wiping dirt off the sonosurg curved scissors, hf, pistol grip,5mm x 34cm during a laparoscopic cholecystectomy.The procedure was completed with a similar device.There were no reports of patient harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information obtained from the customer regarding the reported event and on the legal manufacturer's final investigation.A review of the device history record (dhr), found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of this phenomenon could not be identified.The likely cause was the following: 1.Only the distal end of the probe was pushed forcefully against the tissue while the output was activating, or the output was activating while twisting the scissors with grasping the tissue.This caused the probe to have cracks.2.A force was applied to the distal the probe during cleaning, causing the probe to break at the cracks.The instruction for use manual contains the necessary and enough information to handle the device as follows: "activate the probe only when tissue has been grasped.Do not apply the probe to the tissue with strong force, twist the probe while grasping tissue, or pull the probe up grasping tissue.Otherwise, the probe may be damaged by interference with the internal parts or load increases, which could result in the tip breaking and dropping it inside the body cavity.Do not activate ultrasonic output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such a metal clips or other instruments (e.G., metal clips, uterine manipulators, or other instruments).Do not apply only the probe to the tissue with strong force or do not twist or pull the probe up while grasping the tissue.Otherwise, the probe may become too hot and break before the generator gives a warning lamp/sound." olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the dirt reported was scorch marks in the surface of the device that fell outside the patient's body during use.There was no delay to the procedure due to this event.It was further reported the device output was checked before use and was normal.There are no additional details regarding the patient or event.
|
| |
|