| Model Number |
25CAVGJ-514 00 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022 a 25mm masters series heart coated aortic valved graft was selected for an implant.During the procedure, the holder handles was fully inserted into the orifice at a 90 degree angle.The physician rotated the valve with the valve rotator, no resistance was felt.The leaflet fell off and was completely removed from the patient.Device was exchange with a new 25mm masters series heart coated aortic valved graft that was implanted.The patient remains hemodynamically stable throughout the procedure.The patient's surgery time was extended by more than an hour, and the cardiopulmonary bypass machine worked for more than an hour, however no injury was reported.The patient is currently begin observed in the intensive care unit.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the leaflet being dislodged and fractured was confirmed.The investigation found that one of the leaflets was dislodged from the orifice.The dislodged leaflet was fractured into multiple pieces and was returned back.The valve orifice was not fractured.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There was no evidence of the material defect in the carbon coating that may have caused or contributed to the dislodged leaflet.Based on the information received, the root cause of the reported incident could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|