MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES Back to Search Results
Model Number M006630222B1
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch report is being filed based on the analysis of the returned guidewire.When the actual device was received on september 6, 2022, the evaluation revealed the following: as received, the specimen consisted of one-1 each hydro gw std s 150-035; returned loaded within dispenser assembly and single bagged within "zip-lock" style poly pouch.The specimen presented an overall length of 150.70cm and a finished diameter of.02810" to.03155".A gage bushing certified to be.0350" passed over the length of the specimen with no more effort than the mass of the bushing sliding down the wire shaft unaided, except by gravity.Note all diameter measurements were acquired with a non-contact, toolmakers scope after allowing the specimen device to become dry after hydrating the specimen device with blood-bank saline.After wetting with blood-bank saline, the specimen was subjected to visual and tactile examination.The specimen coating appeared visually and tactilely consistent when examined at 1x - 18 inches, unaided, wet.The specimen presented a ductile/tensile overload fracture of the polymer jacket with material removal exposing the distal 0.40cm of the metallic core wire; the polymer jacket material distal of the fracture was not returned with the specimen.Except where noted, the specimen device appeared visually and dimensionally correct.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted, the specimen presented a ductile/tensile overload fracture of the polymer jacket with material removal exposing the metallic core wire; the polymer jacket material distal of the fracture was not returned with the specimen.The polymer jacket damage appears consistent with attempting to remove the wire from the dispenser assembly without proper hydration.As indicated in the directions for use (dfu), "the surface of the zipwire hydrophilic guidewire is not lubricious unless it is wet.Before removing the zipwire hydrophilic guidewire from the protective hoop (by its distal end), inject physiological saline solution into the hub end of the holder in order to completely wet the surface of the wire." as noted in the dfu operational instructions, "to activate hydrophilic coating: before removing the zipwire hydrophilic guidewire from the protective hoop (by its distal end), inject physiological saline solution into the hub end of the holder in order to completely wet the surface of the wire." based on the information provided and the evidence presented, it appeared that handling factors and/or procedural factors have contributed to the event as reported.If there is any further relevant information provided, a follow up medwatch report will be submitted.
 
Event or Problem Description
It was reported that the device could not be used during the preparation.The device is expected to be returned before july 1st.There were no patient complications reported.What was the patient condition following procedure? stable event resolved? yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
UROLOGICAL CATHETER AND ACCESSORIES
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
victoria foss
340 lake hazeltine drive
chaska, MN 55318
9526418506
MDR Report Key15471670
Report Number2126666-2022-00036
Device Sequence Number11525720
Product Code EYA
UDI-Device Identifier08714729755326
UDI-Public08714729755326
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberM006630222B1
Device Catalogue Number630-222B
Device Lot Number6378791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Initial Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-