| Model Number |
8884721255 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that the proximal site of the gastric tube y-port cracked easily and result in leakage during feeding.On 19sep2022 it was stated that normally user replaces the device every three months.For this incident, the user needed to replace the device after one month of usage.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.No photos or samples were returned for evaluation.An investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.Based on the available information and since a root cause could not be determined, no action plan is required at this point.We will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.
|
| |
|