ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1070-060
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion in the mid left iliac artery with 80% stenosis, heavy calcification and mild tortuosity.The 7.0x60mm armada 35 balloon dilatation catheter (bdc) had no resistance during advancement, however, burst during the first inflation at 4 atmospheres (atms) in a circular form.The armada 35 bdc was soaked and prepared (air aspiration) outside the anatomy prior to use.There was resistance during removal when trying to withdraw the delivery system of the balloon as a unit.Upon removal, the balloon and delivery catheter were noted to be separated and surgical arteriotomy was performed to remove the proximal part of the balloon and delivery catheter.There was no adverse patient sequela.There was a delay of 1 hour in the procedure due to the surgery.No additional information was provided.
 
Additional Manufacturer Narrative
A visual and scanning electron microscopy (sem) inspection/analysis were performed on the returned device.The reported balloon rupture and separation were confirmed.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture, difficulty removing the device and separation appear to be related to circumstances of the procedure.Balloon material ruptures can be affected by numerous factors including, but not limited to, balloon damage during processing of the balloon material, materials, inflation technique, and interactions with other devices or lesion calcification.Based on the reported information and analysis of the returned device, it is likely that the balloon rupture was the result of interaction with the heavily calcified lesion.Additionally, during removal there was resistance as the balloon interacted with the sheath causing the balloon and inner member material to separate.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15473203
Report Number2024168-2022-09965
Device Sequence Number2873466
Product Code LIT
UDI-Device Identifier08717648154799
UDI-Public08717648154799
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberB1070-060
Device Catalogue NumberB1070-060
Device Lot Number20126G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date11/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-