| Model Number |
11-6401 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The explanted locking screws and other hardware were returned for evaluation.The results of the evaluation are inprocess.
|
| |
|
Event or Problem Description
|
|
At an unspecified time post-operatively, a surgical procedure was performed to replace locking screws that had dissociated from the screw heads in a multi-level cervico-thoracic construct.Additional hardware was removed and replaced at that time.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned were two (2) locking screws and in addition the other hardware that was removed during the procedure including two (2) screw-yoke assemblies and two (2) connectors with locking screws.The spinal rods used in this pedicle screw construct were not returned.The returned devices were evaluated.Visual inspection of the all devices found indicators consistent with use in a pedicle screw construct.One locking screw and one screw assembly displayed deformation consistent with the reported event.Review of production records did not identified issues related to production or inspection that may be associated to the reported event.Based on the evaluation findings, the information available and as not all devices used in the construct were available for evaluation, the exact root cause of this event could not be defiantly established.
|
| |
|