Model Number PRT-23
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Arrhythmia (1721); Cardiac Arrest (1762)
Date of Event 08/24/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
This event is being conservatively reported as the cause of the patient¿s death was due to perforation caused by non-abbott devices (ptfe guidewire and al-1 catheter).It was reported that on (b)(6) 2022, during the procedure a non-abbott catheter (al-1 catheter) was initially found to be in a bad position.A pigtail catheter was used to reposition the guidewire.Delivery of the 23mm portico valve was attempted 3 times and the valve was fully recaptured to reposition for optimal depth.There was some difficulty with maintaining optimal position of the portico valve due to mild calcification and arrhythmia, the patient was placed on a temporary pacemaker at 120 beats per minute, which helped successfully implant the valve.Aortography was performed and there was minimal aortic insufficiency that decreased after a few minutes.At the time of release the patient fell into cardiac arrest and resuscitation maneuvers were performed.The patient had become hypotensive despite medication and resuscitation measures.The patient was stabilized and the cause was reported to be pericardial effusion with tamponade due to posterior perforation as seen by transthoracic echocardiogram.Attempt to drain the pericardial effusion was unsuccessful and the patient was sent to surgery.A thoracotomy was performed, attempts to close the perforation was done using a patch.Due to capillary fragility the bleeding could not be controlled.The patient was proclaimed deceased.The patient was treated with multiple medications, direct cardiac massage, various blood products.It was thought that the perforation was due to the crossing of the native valve with the non-abbott catheter (al-1 catheter) at the time of non-abbott guidewire (ptfe guidewire) exchange.The cause of death was noted as hemorrhagic hypovolemic shock due to the perforation.That patient's condition prior to the procedure was reported as very fragile, high risk for surgery and critical aortic stenosis.
 
Additional Manufacturer Narrative
An event of cardiac perforation and patient death due to hemorrhagic hypovolemic shock due perforation during the portico valve implant procedure was reported.It was also reported that the patient had high fragility prior to the procedure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the perforation was thought to be caused by the non-abbot catheter and guidewire during crossing of the native valve at the time of the guidewire exchange.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTICO
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15473923
Report Number2135147-2022-01333
Device Sequence Number4323053
Product Code NPT
UDI-Device Identifier05415067010728
UDI-Public05415067010728
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/26/2023
Device Model NumberPRT-23
Device Catalogue NumberPRT-23
Device Lot Number8527669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FLEXNAV DELIVERY SYSTEM SM, FNAV-DS-SM, LOT8311221.; FLEXNAV LOADING SYSTEM SM, FNAV-LS-SM, LOT 8219874.
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-