• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MASTERS SERIES HEART COATED AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MASTERS SERIES HEART COATED AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23CAVGJ-514 00
Medical Device Problem Codes Incomplete Coaptation (2507); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Urinary Tract Infection (2120); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2002, a 23mm masters valve was implanted.Patient maintains inr around 2.0 and has been compliant.On (b)(6) 2022, decreased leaflet mobility was observed and medical treatment was administered.The patient is planned for a procedure to "either clean out valve or replace" but this has not been scheduled.No patient consequences were reported.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of decreased leaflet mobility was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.D6: implant date corrected h6: code 2120 (urinary tract infection) was removed.Code 4582 (no clinical signs, symptoms or conditions) was added.H6: code 2993 (adverse event without identified device or use problem) was removed.Code 2507 (incomplete coaptation) was added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES HEART COATED AORTIC VALVED GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
177 county road b east
woodridge MN 55117
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15474028
Report Number2135147-2022-01334
Device Sequence Number14161232
Product Code LWQ
UDI-Device Identifier05414734006149
UDI-Public05414734006149
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number23CAVGJ-514 00
Device Catalogue Number23CAVGJ-514 00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR