Model Number T3ST411
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the doctor extracted the remaining maxillary teeth and placed a total of (6) implants.However, implant site #9 became fractured at the labial cortical plate during placement of the implant.The implant was not stable so it was removed because the doctor stated that the other (5) implants placed were stable and in good bone which should be adequate to retain the prosthesis.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.One t3 pro non-platform switched tapered implant 4mm (d) x 11.5mm (l) (t3st411) was returned for investigation.No damage identified or signs of malfunction that would contribute to the event.Measurements match drawing.Based on the evaluation and dimensional analysis, device malfunction has not occurred.Additionally, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported devices that did or could cause or contribute to the reported event.Monthly post market trending review identified no adverse events or actionable trends for the reported event or devices.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimvie.Dhr review for the lot number ((2022030713) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.Products was conforming at the time it left zimvie.Lot was inspected and passed all acceptance criteria by qa.All sterilization activities were conducted with no nonconformities per sterilization record (op0210).Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming products within specifications.Complaint history review was performed for the reported lot number (2022030713) for similar events.And no other complaint were identified.The devices could not be functionally tested for the reported failure (bone fracture).The reported event could not be recreated.The complaint is not related to the functional performance of the devices.However, based on the investigation, risk review and ifu, the most likely cause determined from the investigation is clinician places implant in a patient with patient factors (e.G.Poor hygiene, periodontal issues, pre-existing medical conditions) or inadequate treatment planning.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at this time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3 PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 4MM (D) X 11.5MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15474234
Report Number0001038806-2022-01474
Device Sequence Number16873558
Product Code DZE
UDI-Device Identifier00844868046929
UDI-Public(01)00844868046929(17)270422(10)2022030713
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K213672
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberT3ST411
Device Catalogue NumberT3ST411
Device Lot Number2022030713
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-