| Model Number |
NEURO ZTI EVO |
| Medical Device Problem Codes |
Failure to Read Input Signal (1581); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While this device has gained initial pma approval, it has not been commercialized, imported or sold in the us and no implantations have been performed as of the date of this report.The implant neuro zti evo was not explanted as of the date of this report.Currently, this event is not a serious injury.Follow-up reports will be submitted once explant is scheduled and device is received.The subject device is part of the voluntary field corrective action initiated for neuro zti on october 14th 2021 (international recall #211014).
|
| |
|
Event or Problem Description
|
|
In vivo analysis was performed on may 18th, 2022 by oticon medical representative.Oticon medical representative reported the impossibility to connect with the neuro zti evo implant, serial number (b)(4) despite changes in cable, antenna, magnet and sound processor.Electrical artefacts measurements and impedance measurements could not be performed due to the inability of connection with the device.
|
| |
|
Search Alerts/Recalls
|