• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD TRIAD SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD TRIAD SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1108
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A facility reported that the lock on the mayfield triad skull clamp (a1108) was too loose.The facility has not reported if there was patient involvement, injury, or surgical delay relative to this event.
 
Additional Manufacturer Narrative
The mayfield triad skull clamp (a1108) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: on inspection of the unit, it was discovered that the lock had rotation and lateral movement with a residue buildup present.New components were added to replace worn internal parts.Root cause: the reported complaint was confirmed.Unit was in used condition with visible wear to internal components.These components were replaced to restore proper functionality and preventative maintenance and cleaning performed.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD TRIAD SKULL CLAMP
Common Device Name
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15474472
Report Number3004608878-2022-00193
Device Sequence Number4323099
Product Code HBL
UDI-Device Identifier10381780253518
UDI-Public10381780253518
Combination Product (Y/N)N
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA1108
Device Catalogue NumberA1108
Device Lot Number184
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Initial Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-