Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a hip revision case using a g7 acetabular cup, the surgeon could not lock the insert to the shell.He tried again with a new insert but the insert did not lock in place again.Finally, he used a dual mobility insert and head.No adverse events were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign: greece.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event; please see associated report: 0001825034-2022-02196.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified there is damage to the locking ring feature of the device and indentations on the inside sphere.It is unknown if the damage occurred prior or during the implantation attempts.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL ARCOMXL LNR 36MM H
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15474574
Report Number0001825034-2022-02195
Device Sequence Number2892585
Product Code PBI
UDI-Device Identifier00880304526112
UDI-Public(01)00880304526112(17)221221(10)6197850
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/21/2022
Device Model NumberN/A
Device Catalogue Number010000821
Device Lot Number6197850
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date09/23/2022
Supplement Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CAT#: 010000821 G7 LINER LOT#: 6263697; CAT#: 110010269 G7 SHELL LOT#: 7267689
Patient SexUnknown
-
-