• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24634
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the blade came loose.A 6.00mmx2.0cmx135cm peripheral cutting balloon was selected for use.During the procedure, it was noted that the blade came loose when they were pulling it out of the sheath at the end of the case.No complications reported and no patient was harmed.
 
Event or Problem Description
It was reported that the blade came loose.A 6.00mmx2.0cmx135cm peripheral cutting balloon was selected for use.During the procedure, it was noted that the blade came loose when they were pulling it out of the sheath at the end of the case.No complications reported and no patient was harmed.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.A visual examination of the returned device identified that approximately 12mm of a blade were noted to have lifted on one of the blades.All other blades were present and fully bonded to the balloon surface.The customers introducer sheath was not returned with the device.A visual examination of the returned device identified that the balloon had been inflated and was not refolded.It is not known when the balloon was subjected to positive pressure.The balloon of the device was visually examined, and no issues were noted that may have potentially contributed to the complaint incident.No issues were noted with the tip section of the device.A visual examination found no issue with the markerbands.A visual and tactile examination identified no kinks or damage to the shaft of the returned device.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15475374
Report Number2124215-2022-36656
Device Sequence Number4323490
Product Code PNO
UDI-Device Identifier08714729813729
UDI-Public08714729813729
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K151253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2023
Device Model Number24634
Device Catalogue Number24634
Device Lot Number0028149987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-