• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE Back to Search Results
Model Number M00561311
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a captivator large oval med stiff snare was used in the descending colon during a removal of 22mm polyps procedure performed on (b)(6) 2022.During the procedure and inside the patient, the device could not remove the polyposis of the patient.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.The patient condition following the procedure was reported to be stable.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Block h6: problem code a050702 captures the reportable event of snare loop cutting problems.Block h10: (product investigation) one captivator snare was received for analysis.Visual analysis of the device and the active cord found no device problems.Functional evaluation of the returned device found that when the device was connected to the 10 inch loop fixture, it contracted and extended well.Continuity test and dimensional test were also performed and the device's electrical resistance were within specification.No other device problems were noted.The reported event of "loop failure to cut" could not be confirmed since the device cannot be functionally evaluated with respect to the anatomical/procedural factors encountered during the procedure.The product record review confirmed that this was not a new failure type and the risk was anticipated.There was no evidence of a manufacturing problem, design or user problem which could have caused the complaint.Device analysis found no problems with the device during visual, functional, and continuity test.Based on the analysis of the returned device and the information available, the code selected as the most probable cause is no problem detected.This code was selected since the reported event could not be confirmed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator large oval med stiff snare was used in the descending colon during a removal of 22mm polyps procedure performed on (b)(6) 2022.During the procedure and inside the patient, the device could not remove the polyposis of the patient.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.The patient condition following the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15476261
Report Number3005099803-2022-05400
Device Sequence Number6005799
Product Code FDI
UDI-Device Identifier08714729019251
UDI-Public08714729019251
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K131700
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberM00561311
Device Catalogue Number6131
Device Lot Number0028479119
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
Patient Weight68 KG
-
-