| Model Number |
FT18EN50NGF061N |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One sample was returned.A visual inspection of the device revealed no defects.Air was applied to the devices and no leaks were detected.The cuff was filled with 8cc of water and allowed to sit for 2 hours with no signs of leaking.No lot number was provided; therefore, device history record review could not be completed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was brand new out of box and the cuff was leaking.No patient injury was reported.
|
| |
|