(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.In this case, it was reported that the balloon was inflated to 8 atmospheres.It should be noted that the armada 35 instruction for use (ifu) states: ¿inflation in excess of the rated burst pressure may cause the balloon to rupture.Use of a pressure monitoring device is recommended.¿ the rated burst pressure for this device is 7 atmospheres as indicated on the product label.It could not be determined if inflating the balloon slightly over the rated burst pressure contributed to the rupture.The investigation determined the reported balloon rupture appears to be related to operational context.Balloon material ruptures can be affected by numerous factors including, but not limited to, balloon damage during processing of the balloon material, materials, inflation technique, and interactions with other devices or lesion calcification.In this case, it is likely that the balloon rupture occurred due to interaction with the moderately calcified anatomy.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat a moderately calcified, 70% stenosed lesion in the left superior femoral artery (sfa).A 14x80mm armada 35 balloon dilatation catheter (bdc) was advanced to the target lesion and inflated one time to 8 atmospheres (atm) when the balloon ruptured.The bdc was removed, and a new balloon was used to successfully complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|