• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D010
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time, the product has not been returned.If the product is returned and analysis will be performed, a supplemental medwatch will be filed if there is any further relevant information from that review.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) showed eleven months of battery remaining then it tripped elective replacement indicator (eri).Data analysis indicated that the device was depleting normally and eri was declared due to two hv charges.This is a secondary mechanism for setting eri and is not tracked by the longevity calculation algorithm.This device is an outlier in charging performance but is not malfunctioning.Additional information indicated that the device was explanted due to premature battery depletion (pbd).A new device was implanted.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) showed eleven months of battery remaining then it tripped elective replacement indicator (eri).Data analysis indicated that the device was depleting normally and eri was declared due to two hv charges.This is a secondary mechanism for setting eri and is not tracked by the longevity calculation algorithm.This device is an outlier in charging performance but is not malfunctioning.Additional information indicated that the device was explanted due to premature battery depletion (pbd).A new device was implanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
At this time, the product has not been returned.If the product is returned and analysis will be performed, a supplemental medwatch will be filed if there is any further relevant information from that review.The returned implantable cardioverter defibrillator (icd) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN MINI ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15477213
Report Number2124215-2022-37786
Device Sequence Number5594860
Product Code LWS
UDI-Device Identifier00802526533785
UDI-Public00802526533785
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/16/2016
Device Model NumberD010
Device Catalogue NumberD010
Device Lot Number102505
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer05/08/2023
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date05/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-