• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TCM1510X3
Medical Device Problem Codes Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Emotional Changes (1831); Unspecified Infection (1930); Inflammation (1932); Muscle Weakness/Atrophy (1967); Neuropathy (1983); Pain (1994); Scar Tissue (2060); Urinary Retention (2119); Hernia (2240); Distress (2329); Discomfort (2330); Abdominal Distention (2601); Constipation (3274); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced pain, adhesion, inflammation, bowel problems, infections, defective device, mesh failure, suffering, mental anguish, emotional distress, discomfort, disability, impairment, and loss of enjoyment of life.Post-operative patient treatment included revision surgery, mesh removal, and mesh revision.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced pain, adhesion, inflammation, bowel problems, infections, defective device, mesh failure, suffering, mental anguish, emotional distress, discomfort, disability, impairment, and loss of enjoyment of life.Post-operative patient treatment included revision surgery, mesh removal, and mesh revision.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced pain, adhesion, inflammation, bowel problems, infections, defective device, mesh failure, suffering, mental anguish, emotional distress, discomfort, disability, impairment, loss of enjoyment of life, gait instability, issues with urination, neuropathic pain of right groin abdominal wall, scarring, urinary retention, constipation, nociceptive pain, tenderness, bulging of inguinal canal, debilitation, hernia recurrence, scar tissue.Post-operative patient treatment included revision surgery, mesh removal, mesh revision, diagnostic lap, pain medication, nerve blocks, trigger point injections, modified mcvay repair, ethibond sutures removed, lap retroperitoneal exploration, neurectomy, hospitalization, foley catheter placement, right groin exploration, hernia repair with mesh, removal of all sutures.
 
Additional Manufacturer Narrative
Additional info: a1, a4, b2, b5, b7, h6 (patient codes).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15477416
Report Number9615742-2022-00842
Device Sequence Number6070934
Product Code FTL
UDI-Device Identifier10884521178861
UDI-Public10884521178861
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K090858
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2014
Device Model NumberTCM1510X3
Device Catalogue NumberTCM1510X3
Device Lot NumberSJI00247
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer02/08/2023
03/22/2023
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date02/09/2023
04/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability; Hospitalization; Required Intervention;
Patient SexMale
Patient Weight94 KG
-
-