As reported, the basket of an ngage nitinol stone extractor would not close.The issue was discovered when the device was unpacked, prior to patient contact.A new basket was used to complete the procedure.No adverse events have been reported as a result of the alleged malfunction.
|
Summary of event: as reported, the basket of an ngage nitinol stone extractor would not close.The issue was discovered when the device was unpacked, prior to patient contact.A new basket was used to complete the procedure.No adverse events have been reported as a result of the alleged malfunction.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation, however, a picture of the distal end of the device was supplied by the customer.The picture showed that 1 of the 3 basket wires had been pulled free from the basket sheaths that normally hold the basket wires in place.It was also shown that the basket assembly had moved distally relative to the basket sheath.It appeared the basket assembly had been subjected to a tensile force that caused the observed damage.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted.All extractors are are inspected for damage and verified to assure the basket opens and closes properly.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: suggested handling instructions for extractors and forceps important: excessive force could damage device.Based on the available information and results of the investigation, cook has concluded that the cause for the damage could not be conclusively determined.Excessive force may have been inadvertently applied to the device; however, no information is known regarding device handling.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|