• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; TIBIAL BASEPLATE, SIZE 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; TIBIAL BASEPLATE, SIZE 3 Back to Search Results
Model Number TB-2203E-01M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Unspecified Infection (1930); Pain (1994)
Date of Event 08/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was experiencing pain and developed an alleged infection and hematomas.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: tibial poly spacer, tibial hinge component, and distal femur axial pin.The following eleos implants remain implanted from the patient's original surgery: distal femur, tibial baseplate, and cemented stem extension.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00260; #3013450937-2022-00261; #3013450937-2022-00262; #3013450937-2022-00263; #3013450937-2022-00264.
 
Event or Problem Description
It was reported that the patient was experiencing pain and developed an alleged infection and hematomas.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: tibial poly spacer, tibial hinge component, and distal femur axial pin.The following eleos implants remain implanted from the patient's original surgery: distal femur, tibial baseplate, and cemented stem extension.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the pain, alleged infection and hematoma could not be determined.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added.G3: date received by manufacturer added.G6: type of report added.H2: follow-up type added.H3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 3221: no findings available.H6: investigation conclusions code updated to 4315: cause not established.H10: additional narratives/data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
TIBIAL BASEPLATE, SIZE 3
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key15477531
Report Number3013450937-2022-00259
Device Sequence Number20232806
Product Code KRO
UDI-Device IdentifierB278TB2203E01M0
UDI-PublicB278TB2203E01M0
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTB-2203E-01M
Device Catalogue NumberTB-2203E-01M
Device Lot Number90733-010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001208E, ELEOS POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N CP3A0123, MICROPORT CUSTOM DEVICE; P/N CP3A0124, MICROPORT CUSTOM DEVICE; P/N KSC01530E, ELEOS STEM EXTENSION; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age66 YR
Patient SexMale
-
-