Model Number 27050G-S
Medical Device Problem Codes Energy Output Problem (1431); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer's complaint reference number for this case is (b)(4) which reports all incidents in one case.Three related complaints to this event (b)(4), (b)(4) and (b)(4).In a case on (b)(6) 2022 with dr.(b)(6), per the staff in the room, 4 loops out of the same box broke right in the middle of the loop every time (this was true for all of them).No harm reported to patient.
 
Additional Manufacturer Narrative
All 3 of the electrodes sent in have the same evaluation findings.The distal loop is broken.The ends of the remaining loop are burnt.The yellow insulation has melted and there is black residue.At the proximal end of the piece, the black coating has separated from the end.The shaft is bent.The 4th loop was used and discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE,CUTTING,STERILE.24FR
Common Device Name
MONOPOLAR SINGLE USE CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm,
GM  
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245
4242188247
MDR Report Key15477677
Report Number9610617-2022-00272
Device Sequence Number17541787
Product Code HIH
UDI-Device Identifier04048551394540
UDI-Public4048551394540
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K882270
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number27050G-S
Device Catalogue Number27050G-S
Device Lot Number37FB6368
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/24/2022
Initial Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-