D4.Additional implant explanted; model #:336-0358 lot #:sm156302 udi#:(b)(4).D5.Health professional h6.Investigation findings:3252 investigation conclusions: 4310, 22.H10: the dhrs for parts explanted were reviewed, and no non-conformances were noted in manufacturing (material, dimensions, and processes) were all within specification.The implants were used for treatment, not diagnosis.Review of explanted devices confirms that distal shaft of 1 screw is fractured, and blocker component at one distal level of the 3-level plate is fractured at both blocker wings.Based on the confirmed fracture of the blocker, it is likely that both the screw and the blocker of the plate experienced significant post-operative loads due to an unstable construct.During the failure investigation, it was noted in follow-up discussions with the surgeon, during revision, mild infection was found.Little to no fusion at the lowest level.This could increase the likelihood that the implant components would continue to see significant post-operative loading without fusion.Labeling review: "warnings/cautions/precautions: without solid bone fusion, this device cannot be expected to support the cervical spine indefinitely and may fail due to bone-metal interface, metal, or bone failure.Based on the fatigue testing results, the physician/surgeon should consider the levels of implantation, patient weight, patient activity level, compliance of the patient, and other patient conditions which may have an impact on the performance and results of this system.No spinal implant can withstand body loads for an indefinite period of time without the support of bone.In the event that successful fusion is not achieved; bending, breakage, loosening, migration and/or disassembly of the device will occur.Potential risks identified with the use of this device, which may require additional surgery, include device component fracture, loss of fixation, non-union, fracture of the vertebrae, and necrosis of bone, neurological injury and vascular or visceral injury.Possible adverse effects: the following complications and adverse reactions have been shown to occur with the use of similar spinal instrumentation.These effects and any other known by the surgeon should be discussed with the patient preoperatively.Initial or delayed loosening, disassembly, bending, dislocation and/or breakage of device components, non-union or pseudoarthrosis, infection.Postoperative management: in the case of delayed union or non-union of bone, the patient must continue to be immobilized in order to prevent bending, dislocation, or breakage of the implant devices.Immobilization should continue until a complete bone fusion mass has developed and been confirmed.".
|