• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON FLEX 45 LONG; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON FLEX 45 LONG; STAPLE, IMPLANTABLE Back to Search Results
Model Number EC45AL
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the device deliver any staples? if yes, were the staples formed properly? if yes, was the staple line complete? did the device cut? if yes, was the cut line complete? if yes, was there any issue with the cut line such as jagged, dull knife, irregular, etc? was there any difficulty opening the device? if yes, how was the device removed from the patient? if yes, did the jaws eventually open? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the echelon endo stapler stuck with the load, requiring a stapler replacement.Did not cause any harm to the patient.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 3/9/2023.D4: batch # 181a41.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ec45al device was returned with no apparent damage and with an ecr45b reload loaded on the device.The reload was received fully fired.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: if the jaws do not automatically open after the anvil release button is pressed, first ensure that the knife is in the retracted position by verifying that the stroke count indicator displays ¿0¿ and knife direction indicator points towards the proximal side of the instrument, or that the knife blade indicator is in the home position.If the stroke count indicator or knife blade indicator is not in the home position, push the red manual knife reverse switch downward to reverse the knife motion and squeeze the firing trigger, completely until it rests on the closing trigger.Press the anvil release button.If the jaws do not open at this point, then gently pull the closing trigger, upward (away from the handle) until both firing and closing triggers return to their original positions.The event described could not be confirmed as the device performed as intended and it opened and closed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number 181a41, and no non-conformances were identified.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 11/30/2023.An analysis of the product could not be performed since a physical sample was not received for evaluation.The manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Upon further review, it was concluded that the device initially received under this complaint could not be confirmed to belong to this event.No device has been returned for this complaint event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX 45 LONG
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
329348013
MDR Report Key15478977
Report Number3005075853-2022-06406
Device Sequence Number5673559
Product Code GDW
UDI-Device Identifier10705036001683
UDI-Public10705036001683
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberEC45AL
Device Catalogue NumberEC45AL
Device Lot NumberV94N69
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer02/08/2023
11/30/2023
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date03/09/2023
11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECR45B
Patient SexUnknown
-
-