Model Number APB-4-10-HX-SS
Medical Device Problem Codes Separation Failure (2547); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that an axium coil did not detach.During a coiling procedure, the coil could not be detached.The detachment attempts were made with an id-i-1 detacher.Manual detachment was not attempted as the doctor preferred to change the coil.No other anomalies were reported.The physician did not try another instant detacher.There was no issue with the instant detacher.The devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or complications were reported.
 
Additional Manufacturer Narrative
H3: the axium prime coil was returned for evaluation.There was no instant detacher returned with the coil.The implant coil was still attached to the pushwire.No damage was found with the implant coil.The coin located against the lumen stop.No bend was found on the pushwire.The break indicator was present and intact.The ai and coupler tubing were also found present and intact.There was no evidence of mechanical or manual detachment attempts to detach the coil.No other anomalies were observed.An in-house instant detacher was used to detach the coil.The implant coil was detached successfully on the 1st attempt.Based on the analysis performed, the axium prime coil was confirmed to have "non-detachment" issue as the pushwire was returned with the implant coil was still attached.However, the root cause could not be determined as the implant coil was detached successfully using an in-house instant detacher on 1st attempt.No damages were found with the returned coil and pushwire.No issues were found during detachment of the coil.There was no non-conformance to specifications identified that led to the reported issue.Since the instant detacher was not returned; any contribution of the instant detacher to the non-detachment issue could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that no anomalies were noted neither during the insertion nor delivery of the coil.It was observed that there was damage to the pushwire near to the manual marker detachment point.The patient was undergoing treatment for an apex left oblong aneurysm located in the a1 segment.The aneurysm size was 5mm x 2mm.The patient's vessel tortuosity was severe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15480225
Report Number2029214-2022-01619
Device Sequence Number5640050
Product Code KRD
UDI-Device Identifier00847536038547
UDI-Public00847536038547
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2024
Device Model NumberAPB-4-10-HX-SS
Device Catalogue NumberAPB-4-10-HX-SS
Device Lot Number221814281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date09/24/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-