| Model Number |
PVF-L |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022, patient was prepared for a standalone aortic valve replacement surgery.After removing the diseased leaflets and calcium the aorta was sized and a large size perceval plus valve was chosen.The valve was prepared and implanted as desired.There was a portion of the aorta that was a tucked a bit below the height of the stent of the valve, so as the tissue were pulling above the stent and the valve was seen to dislodge from the annulus.The valve was removed from the patient and the surgeon requested the valve to be prepared again for implant.On the second implant, the valve was found to be high on the non coronary cusp side of the annulus.It was reported that there was a good size chunk of calcium on the aortomitral edge.Surgeon felt that it was too risky to remove this calcium due to its location.3rd attempt was made to implant the valve but they had same findings, the valve was dislodging at the non- coronary cusp side likely due to this calcium seen.As such, the perceval plus valve was removed and an edwards inspiris valve was implanted instead.The procedure was then completed as desired and patient was doing well.Reportedly, on echo there was some lvot hypertrophy and heavy calcification making it hard to see the annulus perfectly on initial inspection.The annulus was sized as 24 mm on echo.No dilation of root was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturing and material records for the perceval plus heart valve, model #pvf-l , s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model #pvf-l) perceval plus heart valve at the time of manufacture and release.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, d9, g3, g6, h2, h6.The valve was returned to the manufacturer.The valve was slightly stained with blood but in a general fair condition.Based on the visual inspection performed, it was possible to state that the event is not valve related.The conclusion will remain the same.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer attempted to retrieve further information on this event, and the device involved.However, no further information has been provided to date, despite the manufacturer's attempts.The manufacturer has not yet received the device back.As such, further investigation on the device cannot be performed.Since the device has not yet been returned to the manufacturer, further investigation on the device cannot be performed and the definitive root cause of the reported event cannot be established.However, from the document review performed, no manufacturing deficiencies were identified.Furthermore, based on the information received (i.E.,''there was a good size chunk of calcium on the aortomitral edge'' and ''the valve was dislodging at the non- coronary cusp side likely due to this calcium''), it is possible that patient's anatomy was contributed to the reported event.It should be noted that as explained in the perceval plus instructions for use, "a removed perceval plus prosthesis must not be re-implanted, because its integrity is no longer ensured." as such, the decision to re-collapse the valve was made off-label.Should the device be received in the future, a follow up report will be submitted.
|
| |
|